Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance)Text with EEA relevance

Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance)Text with EEA relevance

CHAPTER I INTRODUCTORY PROVISIONS

Section 1 Scope and definitions

Article 1 Subject matter and scope

1.

This Regulation lays down rules concerning the placing on the market, making available on the market or putting into service of in vitro diagnostic medical devices for human use and accessories for such devices in the Union. This Regulation also applies to performance studies concerning such in vitro diagnostic medical devices and accessories conducted in the Union.

2.

For the purposes of this Regulation, in vitro diagnostic medical devices and accessories for in vitro diagnostic medical devices shall hereinafter be referred to as ‘devices’.

3.

This Regulation does not apply to:

  1. products for general laboratory use or research-use only products, unless such products, in view of their characteristics, are specifically intended by their manufacturer to be used for in vitro diagnostic examination;

  2. invasive sampling products or products which are directly applied to the human body for the purpose of obtaining a specimen;

  3. internationally certified reference materials;

  4. materials used for external quality assessment schemes.

4.

Any device which, when placed on the market or put into service, incorporates, as an integral part, a medical device as defined in point 1 of Article 2 of Regulation (EU) 2017/745 shall be governed by that Regulation. The requirements of this Regulation shall apply to the in vitro diagnostic medical device part.

5.

This Regulation is specific Union legislation within the meaning of Article 2(3) of Directive 2014/30/EU.

6.

Devices which are also machinery within the meaning of point (a) of the second paragraph of Article 2 of Directive 2006/42/EC of the European Parliament and of the Council(1) shall, where a hazard relevant under that Directive exists, also meet the essential health and safety requirements set out in Annex I to that Directive to the extent to which those requirements are more specific than the general safety and performance requirements set out in Chapter II of Annex I to this Regulation.

7.

This Regulation shall not affect the application of Directive 2013/59/Euratom.

8.

This Regulation shall not affect the right of a Member State to restrict the use of any specific type of device in relation to aspects not covered by this Regulation.

9.

This Regulation shall not affect national law concerning the organisation, delivery or financing of health services and medical care, such as the requirement that certain devices may only be supplied on a medical prescription, the requirement that only certain health professionals or health care institutions may dispense or use certain devices or that their use be accompanied by specific professional counselling.

10.

Nothing in this Regulation shall restrict the freedom of the press or the freedom of expression in the media in so far as those freedoms are guaranteed in the Union and in the Member States, in particular under Article 11 of the Charter of Fundamental Rights of the European Union.

Article 2 Definitions

For the purposes of this Regulation, the following definitions apply:

  1. ‘medical device’ means ‘medical device’ as defined in point (1) of Article 2 of Regulation (EU) 2017/745;

  2. in vitro diagnostic medical device’ means any medical device which is a reagent, reagent product, calibrator, control material, kit, instrument, apparatus, piece of equipment, software or system, whether used alone or in combination, intended by the manufacturer to be used in vitro for the examination of specimens, including blood and tissue donations, derived from the human body, solely or principally for the purpose of providing information on one or more of the following:

    1. concerning a physiological or pathological process or state;

    2. concerning congenital physical or mental impairments;

    3. concerning the predisposition to a medical condition or a disease;

    4. to determine the safety and compatibility with potential recipients;

    5. to predict treatment response or reactions;

    6. to define or monitoring therapeutic measures.

    Specimen receptacles shall also be deemed to be in vitro diagnostic medical devices;

  3. ‘specimen receptacle’ means a device, whether of a vacuum-type or not, specifically intended by its manufacturer for the primary containment and preservation of specimens derived from the human body for the purpose of in vitro diagnostic examination;

  4. ‘accessory for an in vitro diagnostic medical device’ means an article which, whilst not being itself an in vitro diagnostic medical device, is intended by its manufacturer to be used together with one or several particular in vitro diagnostic medical device(s) to specifically enable the in vitro diagnostic medical device(s) to be used in accordance with its/their intended purpose(s) or to specifically and directly assist the medical functionality of the in vitro diagnostic medical device(s) in terms of its/their intended purpose(s);

  5. ‘device for self-testing’ means any device intended by the manufacturer to be used by lay persons, including devices used for testing services offered to lay persons by means of information society services;

  6. ‘device for near-patient testing’ means any device that is not intended for self-testing but is intended to perform testing outside a laboratory environment, generally near to, or at the side of, the patient by a health professional;

  7. ‘companion diagnostic’ means a device which is essential for the safe and effective use of a corresponding medicinal product to:

    1. identify, before and/or during treatment, patients who are most likely to benefit from the corresponding medicinal product; or

    2. identify, before and/or during treatment, patients likely to be at increased risk of serious adverse reactions as a result of treatment with the corresponding medicinal product;

  8. ‘generic device group’ means a set of devices having the same or similar intended purposes or a commonality of technology allowing them to be classified in a generic manner not reflecting specific characteristics;

  9. ‘single-use device’ means a device that is intended to be used during a single procedure;

  10. ‘falsified device’ means any device with a false presentation of its identity and/or of its source and/or its CE marking certificates or documents relating to CE marking procedures. This definition does not include unintentional non-compliance and is without prejudice to infringements of intellectual property rights;

  11. ‘kit’ means a set of components that are packaged together and intended to be used to perform a specific in vitro diagnostic examination, or a part thereof;

  12. ‘intended purpose’ means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements or as specified by the manufacturer in the performance evaluation;

  13. ‘label’ means the written, printed or graphic information appearing either on the device itself, or on the packaging of each unit or on the packaging of multiple devices;

  14. ‘instructions for use’ means the information provided by the manufacturer to inform the user of a device's intended purpose and proper use and of any precautions to be taken;

  15. ‘Unique Device Identifier’ (‘UDI’) means a series of numeric or alphanumeric characters that is created through internationally accepted device identification and coding standards and that allows unambiguous identification of specific devices on the market;

  16. ‘risk’ means the combination of the probability of occurrence of harm and the severity of that harm;

  17. ‘benefit-risk determination’ means the analysis of all assessments of benefit and risk of possible relevance for the use of the device for the intended purpose, when used in accordance with the intended purpose given by the manufacturer;

  18. ‘compatibility’ is the ability of a device, including software, when used together with one or more other devices in accordance with its intended purpose, to:

    1. perform without losing or compromising the ability to perform as intended, and/or

    2. integrate and/or operate without the need for modification or adaption of any part of the combined devices, and/or

    3. be used together without conflict/interference or adverse reaction;

  19. ‘interoperability’ is the ability of two or more devices, including software, from the same manufacturer or from different manufacturers, to:

    1. exchange information and use the information that has been exchanged for the correct execution of a specified function without changing the content of the data, and/or

    2. communicate with each other, and/or

    3. work together as intended;

  20. ‘making available on the market’ means any supply of a device, other than a device for performance study, for distribution, consumption or use on the Union market in the course of a commercial activity, whether in return for payment or free of charge;

  21. ‘placing on the market’ means the first making available of a device, other than a device for performance study, on the Union market;

  22. ‘putting into service’ means the stage at which a device, other than a device for performance study, has been made available to the final user as being ready for use on the Union market for the first time for its intended purpose;

  23. ‘manufacturer’ means a natural or legal person who manufactures or fully refurbishes a device or has a device designed, manufactured or fully refurbished, and markets that device under its name or trade mark;

  24. ‘fully refurbishing’, for the purposes of the definition of manufacturer, means the complete rebuilding of a device already placed on the market or put into service, or the making of a new device from used devices, to bring it into conformity with this Regulation, combined with the assignment of a new lifetime to the refurbished device;

  25. ‘authorised representative’ means any natural or legal person established within the Union who has received and accepted a written mandate from a manufacturer, located outside the Union, to act on the manufacturer's behalf in relation to specified tasks with regard to the latter's obligations under this Regulation;

  26. ‘importer’ means any natural or legal person established within the Union that places a device from a third country on the Union market;

  27. ‘distributor’ means any natural or legal person in the supply chain, other than the manufacturer or the importer, that makes a device available on the market, up until the point of putting into service;

  28. ‘economic operator’ means a manufacturer, an authorised representative, an importer or a distributor;

  29. ‘health institution’ means an organisation the primary purpose of which is the care or treatment of patients or the promotion of public health;

  30. ‘user’ means any healthcare professional or lay person who uses a device;

  31. ‘lay person’ means an individual who does not have formal education in a relevant field of healthcare or medical discipline;

  32. ‘conformity assessment’ means the process demonstrating whether the requirements of this Regulation relating to a device have been fulfilled;

  33. ‘conformity assessment body’ means a body that performs third-party conformity assessment activities including calibration, testing, certification and inspection;

  34. ‘notified body’ means a conformity assessment body designated in accordance with this Regulation;

  35. ‘CE marking of conformity’ or ‘CE marking’ means a marking by which a manufacturer indicates that a device is in conformity with the applicable requirements set out in this Regulation and other applicable Union harmonisation legislation providing for its affixing;

  36. ‘clinical evidence’ means clinical data and performance evaluation results, pertaining to a device of a sufficient amount and quality to allow a qualified assessment of whether the device is safe and achieves the intended clinical benefit(s), when used as intended by the manufacturer;

  37. ‘clinical benefit’ means the positive impact of a device related to its function, such as that of screening, monitoring, diagnosis or aid to diagnosis of patients, or a positive impact on patient management or public health;

  38. ‘scientific validity of an analyte’ means the association of an analyte with a clinical condition or a physiological state;

  39. ‘performance of a device’ means the ability of a device to achieve its intended purpose as claimed by the manufacturer. It consists of the analytical and, where applicable, the clinical performance supporting that intended purpose;

  40. ‘analytical performance’ means the ability of a device to correctly detect or measure a particular analyte;

  41. ‘clinical performance’ means the ability of a device to yield results that are correlated with a particular clinical condition or a physiological or pathological process or state in accordance with the target population and intended user;

  42. ‘performance study’ means a study undertaken to establish or confirm the analytical or clinical performance of a device;

  43. ‘performance study plan’ means a document that describes the rationale, objectives, design methodology, monitoring, statistical considerations, organisation and conduct of a performance study;

  44. ‘performance evaluation’ means an assessment and analysis of data to establish or verify the scientific validity, the analytical and, where applicable, the clinical performance of a device;

  45. ‘device for performance study’ means a device intended by the manufacturer to be used in a performance study.

    A device intended to be used for research purposes, without any medical objective, shall not be deemed to be a device for performance study;

  46. ‘interventional clinical performance study’ means a clinical performance study where the test results may influence patient management decisions and/or may be used to guide treatment;

  47. ‘subject’ means an individual who participates in a performance study and whose specimen(s) undergo in vitro examination by a device for performance study and/or by a device used for control purposes;

  48. ‘investigator’ means an individual responsible for the conduct of a performance study at a performance study site;

  49. ‘diagnostic specificity’ means the ability of a device to recognise the absence of a target marker associated with a particular disease or condition;

  50. ‘diagnostic sensitivity’ means the ability of a device to identify the presence of a target marker associated with a particular disease or condition;

  51. ‘predictive value’ means the probability that a person with a positive device test result has a given condition under investigation, or that a person with a negative device test result does not have a given condition;

  52. ‘positive predictive value’ means the ability of a device to separate true positive results from false positive results for a given attribute in a given population;

  53. ‘negative predictive value’ means the ability of a device to separate true negative results from false negative results for a given attribute in a given population;

  54. ‘likelihood ratio’ means the likelihood of a given result arising in an individual with the target clinical condition or physiological state compared to the likelihood of the same result arising in an individual without that clinical condition or physiological state;

  55. ‘calibrator’ means a measurement reference material used in the calibration of a device;

  56. ‘control material’ means a substance, material or article intended by its manufacturer to be used to verify the performance characteristics of a device;

  57. ‘sponsor’ means any individual, company, institution or organisation which takes responsibility for the initiation, for the management and setting up of the financing of the performance study;

  58. ‘informed consent’ means a subject's free and voluntary expression of his or her willingness to participate in a particular performance study, after having been informed of all aspects of the performance study that are relevant to the subject's decision to participate or, in the case of minors and of incapacitated subjects, an authorisation or agreement from their legally designated representative to include them in the performance study;

  59. ‘ethics committee’ means an independent body established in a Member State in accordance with the law of that Member State and empowered to give opinions for the purposes of this Regulation, taking into account the views of laypersons, in particular patients or patients' organisations;

  60. ‘adverse event’ means any untoward medical occurrence, inappropriate patient management decision, unintended disease or injury or any untoward clinical signs, including an abnormal laboratory finding, in subjects, users or other persons, in the context of a performance study, whether or not related to the device for performance study;

  61. ‘serious adverse event’ means any adverse event that led to any of the following:

    1. a patient management decision resulting in death or an imminent life-threatening situation for the individual being tested, or in the death of the individual's offspring,

    2. death,

    3. serious deterioration in the health of the individual being tested or the recipient of tested donations or materials, that resulted in any of the following:

      1. life-threatening illness or injury,

      2. permanent impairment of a body structure or a body function,

      3. hospitalisation or prolongation of patient hospitalisation,

      4. medical or surgical intervention to prevent life-threatening illness or injury or permanent impairment to a body structure or a body function,

      5. chronic disease,

    4. foetal distress, foetal death or a congenital physical or mental impairment or birth defect;

  62. ‘device deficiency’ means any inadequacy in the identity, quality, durability, reliability, safety or performance of a device for performance study, including malfunction, use errors or inadequacy in information supplied by the manufacturer;

  63. ‘post-market surveillance’ means all activities carried out by manufacturers in cooperation with other economic operators to institute and keep up to date a systematic procedure to proactively collect and review experience gained from devices they place on the market, make available on the market or put into service for the purpose of identifying any need to immediately apply any necessary corrective or preventive actions;

  64. ‘market surveillance’ means the activities carried out and measures taken by public authorities to check and ensure that devices comply with the requirements set out in the relevant Union harmonisation legislation and do not endanger health, safety or any other aspect of public interest protection;

  65. ‘recall’ means any measure aimed at achieving the return of a device that has already been made available to the end user;

  66. ‘withdrawal’ means any measure aimed at preventing a device in the supply chain from being further made available on the market;

  67. ‘incident’ means any malfunction or deterioration in the characteristics or performance of a device made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any harm as a consequence of a medical decision, action taken or not taken on the basis of information or result(s) provided by the device;

  68. ‘serious incident’ means any incident that directly or indirectly led, might have led or might lead to any of the following:

    1. the death of a patient, user or other person,

    2. the temporary or permanent serious deterioration of a patient's, user's or other person's state of health,

    3. a serious public health threat;

  69. ‘serious public health threat’ means an event which could result in imminent risk of death, serious deterioration in a person's state of health, or serious illness, that may require prompt remedial action, and that may cause significant morbidity or mortality in humans, or that is unusual or unexpected for the given place and time;

  70. ‘corrective action’ means action taken to eliminate the cause of a potential or actual non-conformity or other undesirable situation;

  71. ‘field safety corrective action’ means corrective action taken by a manufacturer for technical or medical reasons to prevent or reduce the risk of a serious incident in relation to a device made available on the market;

  72. ‘field safety notice’ means a communication sent by a manufacturer to users or customers in relation to a field safety corrective action;

  73. ‘harmonised standard’ means a European standard as defined in point (1)(c) of Article 2 of Regulation (EU) No 1025/2012;

  74. ‘common specifications’ (CS) means a set of technical and/or clinical requirements, other than a standard, that provides a means of complying with the legal obligations applicable to a device, process or system.

Section 2 Regulatory status of products and counselling

Article 3 Regulatory status of products

Article 4 Genetic information, counselling and informed consent

CHAPTER II MAKING AVAILABLE ON THE MARKET AND PUTTING INTO SERVICE OF DEVICES, OBLIGATIONS OF ECONOMIC OPERATORS, CE MARKING, FREE MOVEMENT

Article 5 Placing on the market and putting into service

Article 6 Distance sales

Article 7 Claims

Article 8 Use of harmonised standards

Article 9 Common specifications

Article 10 General obligations of manufacturers

Article 11 Authorised representative

Article 12 Change of authorised representative

Article 13 General obligations of importers

Article 14 General obligations of distributors

Article 15 Person responsible for regulatory compliance

Article 16 Cases in which obligations of manufacturers apply to importers, distributors or other persons

Article 17 EU declaration of conformity

Article 18 CE marking of conformity

Article 19 Devices for special purposes

Article 20 Parts and components

Article 21 Free movement

CHAPTER III IDENTIFICATION AND TRACEABILITY OF DEVICES, REGISTRATION OF DEVICES AND OF ECONOMIC OPERATORS, SUMMARY OF SAFETY AND CLINICAL PERFORMANCE, EUROPEAN DATABASE ON MEDICAL DEVICES

Article 22 Identification within the supply chain

Article 23 Medical devices nomenclature

Article 24 Unique Device Identification system

Article 25 UDI database

Article 26 Registration of devices

Article 27 Electronic system for registration of economic operators

Article 28 Registration of manufacturers, authorised representatives and importers

Article 29 Summary of safety and performance

Article 30 European database on medical devices

CHAPTER IV NOTIFIED BODIES

Article 31 Authorities responsible for notified bodies

Article 32 Requirements relating to notified bodies

Article 33 Subsidiaries and subcontracting

Article 34 Application by conformity assessment bodies for designation

Article 35 Assessment of the application

Article 36 Nomination of experts for joint assessment of applications for notification

Article 37 Language requirements

Article 38 Designation and notification procedure

Article 39 Identification number and list of notified bodies

Article 40 Monitoring and re-assessment of notified bodies

Article 41 Review of notified body assessment of technical documentation and performance evaluation documentation

Article 42 Changes to designations and notifications

Article 43 Challenge to the competence of notified bodies

Article 44 Peer review and exchange of experience between authorities responsible for notified bodies

Article 45 Coordination of notified bodies

Article 46 List of standard fees

CHAPTER V CLASSIFICATION AND CONFORMITY ASSESSMENT

Section 1 Classification

Article 47 Classification of devices

Section 2 Conformity assessment

Article 48 Conformity assessment procedures

Article 49 Involvement of notified bodies in conformity assessment procedures

Article 50 Mechanism for scrutiny of conformity assessments of class D devices

Article 51 Certificates of conformity

Article 52 Electronic system on notified bodies and on certificates of conformity

Article 53 Voluntary change of notified body

Article 54 Derogation from the conformity assessment procedures

Article 55 Certificate of free sale

CHAPTER VI CLINICAL EVIDENCE, PERFORMANCE EVALUATION AND PERFORMANCE STUDIES

Article 56 Performance evaluation and clinical evidence

Article 57 General requirements regarding performance studies

Article 58 Additional requirements for certain performance studies

Article 59 Informed consent

Article 60 Performance studies on incapacitated subjects

Article 61 Performance studies on minors

Article 62 Performance studies on pregnant or breastfeeding women

Article 63 Additional national measures

Article 64 Performance studies in emergency situations

Article 65 Damage compensation

Article 66 Application for performance studies

Article 67 Assessment by Member States

Article 68 Conduct of a performance study

Article 69 Electronic system on performance studies

Article 70 Performance studies regarding devices bearing the CE marking

Article 71 Substantial modifications to performance studies

Article 72 Corrective measures to be taken by Member States and information exchange between Member States on performance studies

Article 73 Information from the sponsor at the end of a performance study or in the event of a temporary halt or early termination

Article 74 Coordinated assessment procedure for performance studies

Article 75 Review of the coordinated assessment procedure

Article 76 Recording and reporting of adverse events that occur during performance studies

Article 77 Implementing acts

CHAPTER VII POST-MARKET SURVEILLANCE, VIGILANCE AND MARKET SURVEILLANCE

Section 1 Post-market surveillance

Article 78 Post-market surveillance system of the manufacturer

Article 79 Post-market surveillance plan

Article 80 Post-market surveillance report

Article 81 Periodic safety update report

Section 2 Vigilance

Article 82 Reporting of serious incidents and field safety corrective actions

Article 83 Trend reporting

Article 84 Analysis of serious incidents and field safety corrective actions

Article 85 Analysis of vigilance data

Article 86 Implementing acts

Article 87 Electronic system on vigilance and post-market surveillance

Section 3 Market surveillance

Article 88 Market surveillance activities

Article 89 Evaluation of devices suspected of presenting an unacceptable risk or other non-compliance

Article 90 Procedure for dealing with devices presenting an unacceptable risk to health and safety

Article 91 Procedure for evaluating national measures at Union level

Article 92 Other non-compliance

Article 93 Preventive health protection measures

Article 94 Good administrative practice

Article 95 Electronic system on market surveillance

CHAPTER VIII COOPERATION BETWEEN MEMBER STATES, MEDICAL DEVICE COORDINATION GROUP, EU REFERENCE LABORATORIES AND DEVICE REGISTERS

Article 96 Competent authorities

Article 97 Cooperation

Article 98 Medical Device Coordination Group

Article 99 Tasks of the MDCG

Article 100 The European Union reference laboratories

Article 101 Device registers and databanks

CHAPTER IX CONFIDENTIALITY, DATA PROTECTION, FUNDING AND PENALTIES

Article 102 Confidentiality

Article 103 Data protection

Article 104 Levying of fees

Article 105 Funding of activities related to designation and monitoring of notified bodies

Article 106 Penalties

CHAPTER X FINAL PROVISIONS

Article 107 Committee procedure

Article 108 Exercise of the delegation

Article 109 Separate delegated acts for different delegated powers

Article 110 Transitional provisions

Article 111 Evaluation

Article 112 Repeal

Article 113 Entry into force and date of application

ANNEXES

ANNEX I

ANNEX II

ANNEX III

ANNEX IV

ANNEX V

ANNEX VI

ANNEX VII

ANNEX VIII

ANNEX IX

ANNEX X

ANNEX XI

ANNEX XII

ANNEX XIII

ANNEX XIV

ANNEX XV